Halogen Testing
Midwest Microlab is capable of testing solid or liquid samples for the total percentage of Bromine, Chlorine, Fluorine, and Iodine. Please refer to the individual Bromine (Br), Chlorine (Cl), Fluorine (F), and Iodine (I) element links, to the right, for more specific details associated with each Halogen Analysis & Techniques. Air Sensitive/ Glove Box handling is not available for any Halogen Testing due to exposure of the sample to environment.
Halogen Testing
Bromine Testing
Chlorine Testing
Fluorine Testing
Iodine Testing
Additional Halogen Testing Information
Halogen Interferences & Samples Size
Midwest Microlab is capable of testing solid or liquid samples for the total percentage of Halogens with a range of 0.04% (4,000 ppm) to 100%. Accuracy for Halogen testing is +/- 0.4. Please refer below for known Halogen interferences and Sample Size according to Customer Theory. Each element Halogen Analysis & Techniques can be found with its individual elements page.
Interferences Associated with Halogen Testing:
Bromine interferences include : Al, Cl, Eu, Hg, I, Mg, Na, Ni, Se, Si, Sn and Ti.
Chlorine interferences include: Al, Br, Eu, Hg, I, Mg, Na, Ni, Se,Si, Sn and Ti.
Fluorine interferences include: Ba(OH)2 , Ca(OH)2 , KOH , LiOH , NaOH , Sr(OH)2 , Al, B, Ba, Ca, Fe, K, La, Li, Na, Si and Sr.
Iodine interferences include: Cl, Cu, and Se.
Please note the interferences above cannot be circumvented during analysis. Irregular results can be reported due to an interference within the molecular formula of the sample. Please note irregular results due to interferences cannot be interpreted by the chemists. Midwest Microlab chemists will perform requested analysis and report results as found with the understanding that the submitter understood the interferences associated and how it may affect results. Please indicate all elements or the molecular formula on the submission form. If Bromine and Chlorine are found within the same sample molecular formula, then neither Bromine nor Chlorine can be performed due to both testing using the same indicator.
Suggested Sample Size according to Total Percentage Theory:
Halogen 0-5 % Single 5-20% Single Above 20% Single
Bromine 15 mg 10 mg 5 mg
Chlorine 30 mg 15 mg 10 mg
Fluorine 25 mg 20 mg 15 mg
Iodine 15 mg 10 mg 5 mg
Halogen 0-5 % Duplicate 5-20% Duplicate Above 20% Duplicate
Bromine 30 mg 20 mg 10 mg
Chlorine 60 mg 30 mg 20 mg
Fluorine 50 mg 40 mg 30 mg
Iodine 30 mg 20 mg 10 mg
Halogen 0-5 % Triplicate 5-20% Triplicate Above 20% Triplicate
Bromine 45 mg 30 mg 15 mg
Chlorine 90 mg 45 mg 30 mg
Fluorine 75 mg 60 mg 45 mg
Iodine 45 mg 30 mg 15 mg
Turnaround Time
Standard Turnaround time for the total percentage of Bromine and Chlorine results is within 4 business days. RUSH services are available for an additional fee per sample,per analysis with per single run. Rush results turnaround time for Bromine and Chlorine analysis is within 2-3 business days.
Fluorine and Iodine testing are complex and labor intensive determination. Standard Turnaround time for the total percentage of Fluorine and Iodine results is within 5 business days. RUSH services are available for an additional fee per sample, per analysis with per single run. Rush results turnaround time for Fluorine and Iodine analysis is within 3-4 business days.
Midwest Microlab does NOT handle any samples containing Osmium, Uranium, or Thallium within the molecular formula.Midwest Microlab is able to perform samples with Depleted Uranium but must be clearly indicated on the submission form. Please indicate on the comments section of the submission form if samples have a high toxicity or strong stench. We do not handle any Finished Pharmaceuticals for release or highly volatile liquid samples with boiling points of less than 40 deg C.
Get Started
Connect with
Midwest Microlab
Email Us
info@midwestlab.com
OFfice
7212 North Shadeland Ave
Suite #110
Indianapolis, IN 46250
Office Number: (317) 849-6606
Fax Number: (317) 849-8534
Monday - Friday
9am - 5pm